Medical devices that contact non-intact skin or mucous membranes are considered semi-critical devices and must undergo high-level disinfection (HLD) before use. Studies have identified several potential limitations of UV-C for HLD of semi-critical medical devices, including a lack of data demonstrating that UV-C irradiance can be uniformly applied to complex surfaces that contain grooves, notches and imperfections. This study focused on ultrasound probes as commonly used medical devices to show the distribution of irradiance on these surfaces. An endocavity bi-plane probe and curved array surface probe with typical surface topology were 3D scanned and modelled and an array of UV-C light-emitting diodes (LEDs) irradiating the probe surfaces was simulated (simulated wavelength: 275nm [peak], power output: 50mW). The simulated chamber wall material was equivalent to highly reflective polished aluminum with a defined reflectance of 79% at 275nm. To calculate the cycle time required to achieve HLD on probe surfaces, a minimum effective dosage of 1500mJ/cm2 based on published research was used. The simulated irradiance distribution showed a large difference between the points of highest and lowest irradiance (maximum/minimum ratio: 14.70 for the surface probe and 12.74 for the endocavity probe). In addition, the presence of shadowing effects adjacent to notches or grooves was evident. By applying an effective UV-C dose from the literature, cycle times of up to 25 minutes would be required to achieve HLD in the minimally irradiated areas of the probes used in the simulation. These findings highlight the need to demonstrate the efficacy of UV-C radiation against worst case organisms in the areas of lowest irradiance on medical devices to provide assurance these devices are reliably high level disinfected.
Suggested Citation
LaReine Yeoh & Luke Cogar & Mayes Barak & Lit Yeen Tan & Gavin Spargo & Jon Burdach, 2024.
"UV-C disinfection of ultrasound probes: Challenges of uneven irradiance on complex surfaces,"
PLOS ONE, Public Library of Science, vol. 19(10), pages 1-14, October.
Handle:
RePEc:plo:pone00:0312931
DOI: 10.1371/journal.pone.0312931
Download full text from publisher
Corrections
All material on this site has been provided by the respective publishers and authors. You can help correct errors and omissions. When requesting a correction, please mention this item's handle: RePEc:plo:pone00:0312931. See general information about how to correct material in RePEc.
If you have authored this item and are not yet registered with RePEc, we encourage you to do it here. This allows to link your profile to this item. It also allows you to accept potential citations to this item that we are uncertain about.
We have no bibliographic references for this item. You can help adding them by using this form .
If you know of missing items citing this one, you can help us creating those links by adding the relevant references in the same way as above, for each refering item. If you are a registered author of this item, you may also want to check the "citations" tab in your RePEc Author Service profile, as there may be some citations waiting for confirmation.
For technical questions regarding this item, or to correct its authors, title, abstract, bibliographic or download information, contact: plosone (email available below). General contact details of provider: https://journals.plos.org/plosone/ .
Please note that corrections may take a couple of weeks to filter through
the various RePEc services.