Abstract

Aseptic loosening due to the progressive periprosthetic bone resorption following total hip replacement is a crucial concern, that causes complications and failure of the arthroplasty surgery. The mismatch in stiffness between the hip implant and the surrounding femoral bone is one of the key factors leading to bone density resorption. This paper aimed to investigate the long-term response of the femoral bone after THR using the Silent Hip stem. For this purpose, a validated thermodynamic-based computational model was used to compute the change in bone density before and after THR. This model incorporated essential factors involved in bone remodeling process, such as mechanical loading, and biochemical affinities. The results of the numerical simulations using 3D finite element analysis were analyzed in five zones of interest qualitatively and quantitatively. Bone density predictions showed notable bone resorption in cervical areas, specifically in zone 1 and zone 5 of −18.7% and −14%, respectively. Conversely, bone formation was observed in the greater trochanter area (zone 2) of +25%. Stress shielding seemed to occur at cervical area due to the reduction in the mechanical loading in this region. Based on the quantitative analysis of the bone density distribution throughout the femoral bone, it appears that the Silent Hip stem achieved less bone resorption compared to conventional hip stem designs reported in the literature, which could be used for active patients.

Suggested Citation

  • Zainab Al-Hajaj & Pouria Tavakkoli Avval & Habiba Bougherara, 2023. "Computational prediction of the long-term behavior of the femoral density after THR using the Silent Hip stem," Computer Methods in Biomechanics and Biomedical Engineering, Taylor & Francis Journals, vol. 26(4), pages 373-382, March.
  • Handle: RePEc:taf:gcmbxx:v:26:y:2023:i:4:p:373-382
    DOI: 10.1080/10255842.2022.2064712
    as

    Download full text from publisher

    File URL: http://hdl.handle.net/10.1080/10255842.2022.2064712
    Download Restriction: Access to full text is restricted to subscribers.

    File URL: https://libkey.io/10.1080/10255842.2022.2064712?utm_source=ideas
    LibKey link: if access is restricted and if your library uses this service, LibKey will redirect you to where you can use your library subscription to access this item
    ---><---

    As the access to this document is restricted, you may want to search for a different version of it.

    More about this item

    Statistics

    Access and download statistics

    Corrections

    All material on this site has been provided by the respective publishers and authors. You can help correct errors and omissions. When requesting a correction, please mention this item's handle: RePEc:taf:gcmbxx:v:26:y:2023:i:4:p:373-382. See general information about how to correct material in RePEc.

    If you have authored this item and are not yet registered with RePEc, we encourage you to do it here. This allows to link your profile to this item. It also allows you to accept potential citations to this item that we are uncertain about.

    We have no bibliographic references for this item. You can help adding them by using this form .

    If you know of missing items citing this one, you can help us creating those links by adding the relevant references in the same way as above, for each refering item. If you are a registered author of this item, you may also want to check the "citations" tab in your RePEc Author Service profile, as there may be some citations waiting for confirmation.

    For technical questions regarding this item, or to correct its authors, title, abstract, bibliographic or download information, contact: Chris Longhurst (email available below). General contact details of provider: http://www.tandfonline.com/gcmb .

    Please note that corrections may take a couple of weeks to filter through the various RePEc services.

    IDEAS is a RePEc service. RePEc uses bibliographic data supplied by the respective publishers.